QTM latched instructions#

Instruction

Signature

Execution Time (ns)

Description

set_digital

enable: I, mask: I, fine_delay: I

enable: R, mask: R, fine_delay: R

4


8

Set the output of sequencer/channel to high or low using enable. The parameter mask should always be set to 1.

set_time_ref

—

4

Configure the reference timestamp to the current sequencer time. This timestamp may be used by the acquire_timetags command to determine time differences. This command must be set during or before the command that closes the acquisition window.

set_scope_en

enable: I

enable: R

4

4

Set enable = 1 to trigger scope/trace acquisition by subsequent acquisition instructions.

Note

Adjustment range of fine_delay is [0, 2047] where each number is a multiple of 1/ 128 ns. This fine resolution is limited by hardware resolution of 39 ps (5 steps). Mask should always be set to 1.