With the LINQ-based feedback architecture, the time-tag sequencers can share TTL counts and time-tags in real-time to several Q1 Sequencers within the cluster.

Time-tag LINQ instructions#

Instruction

Signature

Description

fb_llp_tags_id

id: I/R,
duration: I

Configures the tag id attached to low-latency time-tags sent over LINQ and waits duration ns.

Detected time-tags are streamed in real time with 1 ns resolution while the acquisition window remains open. If id = 0, transmission is disabled.

fb_llp_ttls_id

id: I/R,
duration: I

Configures the tag id attached to low-latency TTL counts sent over LINQ and waits duration ns.

The number of detected photon events above the configured threshold is transmitted at the end of the acquisition window. If id = 0, transmission is disabled.

fb_tdc_tags_id

id: I/R,
duration: I

Configures the tag id attached to high-resolution time-tags sent over LINQ and waits duration ns.

Time-tags are generated with approximately 8 ps resolution and are transmitted after additional processing latency. If id = 0, transmission is disabled.

fb_tdc_tdelta_id

id: I/R,
duration: I

Configures the tag id attached to the time difference between two detected events and waits duration ns.

The time delta is computed between a configured reference channel and source channel (which may be the same) with 1/128 ns resolution and ±16.7 ms range. If id = 0, transmission is disabled.