Sequencer

This section will explain how the sequencers of the Pulsar QCM and QRM are controlled. Every sequencer is controlled using the same functions and parameters, which either take the sequencer index as a parameter or indicate which sequencer they operate on based on the index in their name.

Note

As of version 0.5.0 of the Pulsar QRM, new functionality has been added to the acquisition path (e.g. real-time demodulation, (weighed) integration, discretization, averaging, binning). More details about this functionality will be added to the documentation as soon as possible. For now, please have a look at the binned acquisition tutorial to get started.

Overview

The sequencers are split into the sequence processor, AWG and acquisition paths as shown in the figures below. Each sequence processor controls one AWG path and, in case of the Pulsar QRM, one acquisition path. The AWG path and acquisition path are discussed in more detail in section Pulsar. Each sequencer processor is, in turn, split into a classical and real-time pipeline. The classical pipeline is responsible for any classical instructions related to program flow or arithmetic and the real-time pipeline is responsible for real-time instructions that are used to create the experiment timeline.

Pulsar QCM sequencer with AWG path.

Pulsar QCM sequencer with AWG path.

Pulsar QRM sequencer.

Pulsar QRM sequencer with AWG and acquisition paths.

The sequencers are started and stopped by calling the arm_sequencer(), start_sequencer() and stop_sequencer() functions. Once started they will execute the sequence described in the next section.

Sequence

The sequencers are programmed with a sequence using the sequencer#.sequence() function parameter. This parameter expects a sequence in the form of a JSON compatible file that contains the waveform, weight, acquistion and program information. The JSON file is expected to adhere to the following format:

  • waveforms: Indicates that the following waveforms are intended for the AWG path.

    • waveform name: Replace by string containing the waveform name.

      • data: List of floating point values to express the waveform.

      • index: Integer index used by the Q1ASM program to refer to the waveform.

  • weights: Indicates that the following weight functions are intended for the integration units of the acquisition path (only used by the Pulsar QRM).

    • weight name: Replace by string containing the weight name.

      • data: List of floating point values to express the weight.

      • index: Integer index used by the Q1ASM program to refer to the weight.

  • acquisitions: Indicates that the following acquisitions are available for the acquisition path to refer to (only used by the Pulsar QRM).

    • acquisition name: Replace by string containing the acquisition name.

      • num_bins: Number of bins in acquisition.

      • index: Integer index used by the Q1ASM program to refer to the acquisition.

  • program: Single string containing the entire sequence processor Q1ASM program.

Program

The sequence programs are written in the custom Q1ASM assembly language described in the following sections. All sequence processor instructions are executed by the classical pipeline and the real-time instructions are also executed by the real-time pipeline. These latter instructions are intended to control the AWG and acquisition paths in a real-time fashion. Once processed by the classical pipeline they are queued in the real-time pipeline awaiting further execution. A total of 32 instructions can be queued and once the queue is full, the classical part will stall on any further real-time instructions.

Once execution of the real-time instructions by the real-time pipeline is started, care must be taken to not cause an underrun of the queue. An underrun will potentially cause undetermined real-time behaviour and desynchronize any synchronized sequencers. Therefore, when this is detected, the sequencer is completely stopped. A likely cause of underruns is a loop with a very short (i.e. < 24ns) real-time run-time, since the jump of a loop takes some cycles to be execute by the classical pipeline.

Finally, be aware that moving data into a register using an instruction takes a cycle to complete. This means that when an instruction reads from a register that the previous instruction has written to, a nop instruction must be placed in between these consecutive instructions for the value to be correctly read.

The state of the sequencers, including any errors, can be queried through get_sequencer_state().

Instructions

Instructions

Argument 0

Argument 1

Argument 2

Argument 3

Argument 4

Description

Control

illegal
























Instruction that should not
be executed. If it is
executed, the sequencer
will stop with the illegal
instruction flag set.
stop






Instruction that stops the
sequencer.
nop






























No operation instruction,
that does nothing. It is
used to pass a single cycle
in the classic part of the
sequencer without any
operations.

Jumps

jmp


Immediate,
Register,
Label








Jump to the next
instruction indicated by
argument 0.
jge



Register



Immediate



Immediate,
Register,
Label







If argument 0 is greater
or equal to argument 1,
jump to the instruction
indicated by argument 2.
jlt



Register



Immediate



Immediate,
Register,
Label







If argument 0 is less
than argument 1, jump to
the instruction indicated
by argument 2.
loop




Register




Immediate,
Register,
Label














Subtract argument 0 by
one and jump to the
instruction indicated by
argument 1 until
argument 0 reaches zero.

Arithmetic

move

Immediate,
Register
Register




Argument 0 is moved /
copied to argument 1.
not



Immediate,
Register


Register












Bit-wise invert
argument 0
and move the result to
argument 1.
add


Register


Immediate,
Register

Register






Add argument 1 to
argument 0 and move the
result to argument 2.
sub


Register


Immediate,
Register

Register






Subtract argument 1 from
argument 0 and move the
result to argument 2.
and


Register


Immediate,
Register

Register






Bit-wise AND argument 0
and argument 1 and move
the result to argument 2.
or


Register


Immediate,
Register

Register






Bit-wise OR argument 0
and argument 1 and move
the result to argument 2.
xor


Register


Immediate,
Register

Register






Bit-wise XOR argument 0
and argument 1 and move
the result to argument 2.
asl



Register



Immediate,
Register


Register









Bit-wise left-shift
argument 0 by argument 1
number of bits and move
the result to argument 2.
asr



Register



Immediate,
Register


Register









Bit-wise right-shift
argument 0 by argument 1
number of bits and move the
result to argument 2.

Software request

sw_req




Immediate,
Register



















Generate software request
interrupt with argument 0
value being passed as
interrupt argument
(currently not implemented).

Real-time pipeline instructions

set_mrk

Immediate,
Register























































































































Set marker output channels
to argument 0 (bits 0-3),
where the bit index
corresponds to the channel
index for baseband modules.
For QCM-RF module, bit
indices 0 & 1 correspond to
output enable 1 and 2
respectively; indices 2 & 3
correspond to marker outputs
2 and 1 respectively.
For QRM-RF module, bit
indices 0 & 1 correspond to
input 1 and output 1 switches
respectively; indices 2 & 3
correspond to marker outputs
1 and 2 respectively. The
values are OR´ed by that of
other sequencers. The
parameters are cached and
only updated when the
upd_param, play,
acquire or
acquired_weighed
instructions are executed.
set_freq














Immediate,
Register





































































Set the frequency of the NCO
used by the AWG and
acquisition using
argument 0. The frequency
is divided into 4e9 steps
between -500 and 500 MHz and
expressed as an integer
between -2e9 and 2e9. (e.g.
1 MHz=4e6). The frequency
parameter is cached and only
applied when the
upd_param, play,
acquire or
acquired_weighed
instructions are executed.
reset_ph


































































Reset the absolute phase of
the NCO used by the AWG and
acquisition to 0°. This also
resets any relative phase
offsets that were already
statically or dynamically
set. The reset is cached and
only applied when the
upd_param, play,
acquire or
acquired_weighed
instructions are executed.
set_ph














Immediate,
Register





































































Set the relative phase of
the NCO used by the AWG and
acquisition using
argument 0. The phase is
divided into 1e9 steps
between 0° and 360°,
expressed as an integer
between 0 and 1e9 (e.g
45°=125e6). The phase
parameter is cached and only
updated when the
upd_param, play`,
acquire or
acquired_weighed
instructions are executed.
set_ph_delta













Immediate,
Register
































































Set an offset on top of the
relative phase of the NCO
used by the AWG and
acquisition. The offset is
applied on top of the phase
set using set_ph. See
set_ph for more details
regarding the argument. The
phase parameter is cached
and only updated when the
upd_param, play,
acquire or
acquired_weighed
instructions are executed.
set_awg_gain













Immediate,
Register












Immediate,
Register



















































Set AWG gain for path 0
using argument 0 and path
1 using argument 1. Both
gain values are divided in
2**sample path width steps.
The parameters are cached
and only updated when the
upd_param, play,
acquire or
acquired_weighed
instructions are executed.
The arguments are either
all set through immediates
or registers.
set_awg_offs













Immediate,
Register












Immediate,
Register



















































Set AWG gain for path 0
using argument 0 and path
1 using argument 1. Both
offset values are divided
in 2**sample path width
steps. The parameters are
cached and only updated
when the upd_param,
play, acquire or
acquired_weighed
instructions are executed.
The arguments are
either all set through
immediates or registers.
upd_param






Immediate






























Update the marker, phase,
phase offset, gain and
offset parameters set using
their respective
instructions and then wait
for argument 0 number of
nanoseconds.
play












Immediate,
Register











Immediate,
Register











Immediate




































Update the marker, phase,
phase offset, gain and
offset parameters set using
their respective
instructions, start playing
AWG waveforms stored at
indexes argument 0 on
path 0 and argument 1 on
path 1 and finally wait for
argument 2 number of
nanoseconds. The arguments
are either all set through
immediates or registers.
acquire

















Immediate

















Immediate,
Register
















Immediate



















































Update the marker, phase,
phase offset, gain and
offset parameters set using
their respective
instruction, start the
acquisition referred to
using index argument 0 and
store the bin data in bin
index argument 1, finally
wait for argument 2 number
of nanoseconds. Integration
is executed using a square
weight with a preset length
through the associated
QCoDeS parameter. The
arguments are either all
set through immediates or
registers.
acquire_weighed

















Immediate

















Immediate,
Register
















Immediate,
Register
















Immediate,
Register
















Immediate

















Update the marker, phase,
phase offset, gain and
offset parameters set using
their respective
instruction, start the
acquisition referred to
using index argument 0 and
store the bin data in bin
index argument 1, finally
wait for argument 4 number
of nanoseconds. Integration
is executed using weights
stored at indexes
argument 2 for path 0 and
argument 3 for path 1. The
arguments are either all
set through immediates or
registers.
acquire_ttl

















Immediate

















Immediate,
Register
















Immediate

















Immediate


































Update the marker, phase,
phase offset, gain and
offset parameters set using
their respective
instruction, start the
TTL trigger acquisition
referred to using index
argument 0 and store the
bin data in bin index
argument 1, enable the
acquisition by writing 1 to
argument 2, finally wait
for argument 3 number of
nanoseconds. The TTL trigger
acquisition has to be
actively disabled afterwards
by writing 0 to
argument 2.
wait

Immediate,
Register




Wait for argument 0
number of nanoseconds.
wait_trigger



Immediate,
Register














Wait for external trigger
and then wait for
argument 0 number of
nanoseconds.
wait_sync





Immediate,
Register
























Wait for SYNQ to complete
on all connected sequencers
over all connected
instruments and then wait
for argument 0 number of
nanoseconds.

Note

The duration argument for upd_param, play, acquire, acquire_weighed, wait, wait_trigger and wait_sync needs to a be multiple of 4ns. This will be reduced to 1ns in the future.

Arguments

Arguments

Format

Description

Immediate

#

32-bit decimal value (e.g. 1000)

Register

R#

Register address in range 0 to 63 (e.g. R0)

pointing to a 32-bit unsigned integer

Label

@label

Label name string (e.g. @main)

Labels

Any instruction can be preceded by a label. This label can be used as a reference to that specific instruction. In other words, it can be used as a goto-point by any instruction that can alter program flow (i.e. jmp, jge, jlt and loop). The label must be followed by a ‘:’ character and a whitespace before the actual referenced instruction.

Example

This is a simple example of a Q1ASM program. It enables each marker channel output for 1μs and then stops.

      move      1,R0        # Start at marker output channel 0 (move 1 into R0)
      nop                   # Wait a cycle for R0 to be available.

loop: set_mrk   R0          # Set marker output channels to R0
      upd_param 1000        # Update marker output channels and wait 1μs.
      asl       R0,1,R0     # Move to next marker output channel (left-shift R0).
      nop                   # Wait a cycle for R0 to be available.
      jlt       R0,16,@loop # Loop until all 4 marker output channels have been set once.

      set_mrk   0           # Reset marker output channels.
      upd_param 4           # Update marker output channels.
      stop                  # Stop sequencer.

Waveforms

The waveforms are expressed as a list of floating point values in the range of 1.0 to -1.0 with a resolution of one nanosecond per sample. The AWG path uses these waveforms to parametrically generate pulses on its outputs.

Waveform playback is started by the play instructions. Each waveform is paired with an index, which is used by this instruction to refer to the associated waveform. The waveform is then completely played irrespective of further sequence processor instructions, except when the sequence processor issues the playback of another waveform, in which case the waveform will be stopped and the new waveform will start. When waveforms are not played back-to-back, the intermediate time will be filled by samples with a value of zero.

The programmed waveforms can be retrieved using get_waveforms().

Weights

The weights are expressed as a list of floating point values in the range of 1.0 to -1.0 with a resolution of one nanosecond per sample. The integration units in the acquisition path apply (i.e. multiply) these weights during the integration process when the acquisition path is triggered for weighed integration.

Weighed integration is triggered by the acquire_weighed instruction. Each weight is paired with an index, which is used by this instruction to refer to the associated weight. The weight is then played, like the waveforms discussed in the previous section and determines the length of the integration. The weighed integration process continues irrespective of further sequence processor instructions, except when the sequence processor issues another acquisition using the acquire or acquire_weighed instructions, in which case the integration will be stopped, the result will be stored and a new integration will start.

The programmed weights can be retrieved using get_weights().

Acquisitions

Acquisitions are started by the acquire, acquire_weighed or acquire_ttl instructions and will trigger the capture of 16k input samples on both inputs. This mode of operation is called scope mode and will store the raw input samples in a temporary buffer. Every time an acquisition is started, this temporary memory is overwritten, so it is vital to move the samples from the temporary buffer to a more lasting location before the start of the next acquisition. This is done by calling store_scope_acquisition(), which moves the samples into the specified acquisition in the acquisition list of the sequencer, located in the RAM of the instrument. Multiple acquisitions can be stored in this list before being retrieved from the instrument by simply calling get_acquisitions(). Acquisitions are returned as a dictionary of acquisitions. Scope mode data is located under the scope key as lists of floating point values in a range of 1.0 to -1.0 with a resolution of one nanosecond per sample, as well as an indication if the ADC was out-of-range during the measurement.

Note

Before calling store_scope_acquisition(), be sure to call get_sequencer_state() and get_acquisition_state() in that order. This ensures that both the sequencer has finished and that there is an acquisition ready.

The acquisition path also has an averaging function set through the scope_acq_avg_mode_en_path#() parameters. This enables the automatic accumulation of acquisitions, where sample N of acquisition M is automatically accumulated to sample N of acquisition M+1. This happens while the acquisition is still in the temporary buffer, so after the desired number of averaging acquisitions is completed, call store_scope_acquisition() to store the accumulated result in the acquisition list. Once retrieved from the instrument, the accumulated samples will automatically be divided by the number of averages to get the actual averaged acquisition result.

Tip

For debug purposes, the acquisition path can also be triggered using a trigger level, where if the input exceeds this level, an acquisition is started. See the sequencer#.trigger_mode_acq_path#() and sequencer#.trigger_level_acq_path#() parameters for more information.

TTL trigger acquisitions

The acquire_ttl instruction is used to detect TTL triggers on one of the instrument’s inputs when it exceeds a configurable threshold. The input is selected using sequencer#.ttl_acq_input_select(), while the threshold is set using sequencer#.ttl_acq_threshold(). The instruction enables the TTL trigger acquisition path by setting the enable argument to 1 and needs to actively disable the path again by using the same instruction with its enable argument set to 0. While the TTL trigger acquisition path is enabled, real-time integration and discretization is disabled.

acquire_ttl 0,0,1,1000 #Acquire triggers and wait for 1000ns
acquire_ttl 0,0,0,4    #Stop acquiring triggers and wait for 4ns
stop                   #Stop sequencer

When a rising edge is detected on the input that exceeds the threshold while the path is enabled, a TTL trigger is detected and its ADC value at which it was detected is stored in the bin specified by the instruction’s bin index argument. No new TTL triggers will be detected until a new rising edge exceeds the configured threshold.

When multiple triggers are detected while the path is enabled, the storage strategy is determined by the sequencer#.ttl_acq_auto_bin_incr_en() parameter. When sequencer#.ttl_acq_auto_bin_incr_en() is enabled, the bin index is automatically incremented everytime a TTL trigger is detected. This gives the ability to store every ADC value individually and allows you to count the number of TTL triggers by counting the number of valid bins stored at the end of the sequence. Alternatively, when sequencer#.ttl_acq_auto_bin_incr_en() is disabled the same bin is reused for every detected TTL trigger. This allows averaging the ADC values in hardware and allows you to count the number of TTL triggers by looking at the average count of that bin at the end of the sequence.

The resulting acquisition data can be retrieved using get_acquisitions(). The ADC values at which TTL triggers were detected are stored in the integration key as lists of floating point values in a range of 1.0 to -1.0. They replace the integration results of the regular acquisitions.

Continuous waveform mode

The sequencer also supports a continuous waveform mode of operation, where the waveform playback control of sequence processor is completely bypassed and a single waveform is just played back on a loop. This mode can be enabled using the sequencer#.cont_mode_en_awg_path#() parameter and the waveform can be selected using the sequencer#.cont_mode_waveform_idx_awg_path#() parameter. The waveforms used in this mode must be a multiple of four samples long (i.e. 4ns).

When in continuous mode, simply program, arm, start and stop the sequencer using the regular control functions and parameters (i.e. sequencer#.sequence(), arm_sequencer(), start_sequencer() and stop_sequencer()). However, be aware that the sequencer processor can still control parts of the AWG path, like phase, gain and offset, while the sequencer operates in this mode. Therefore, we advise to program the sequence processor with a single stop instruction.

Note

We realise that the current way of controlling this mode is not optimal, so in the near future we will be implementing additional driver support to streamline this mode.